Comparative Ex Vivo Evaluation of Two Electronic Percussive Testing Devices Measuring the Stability of Dental Implants
No Thumbnail Available
Date
2014
Journal Title
Journal ISSN
Volume Title
Publisher
Amer Acad Periodontology
Open Access Color
OpenAIRE Downloads
OpenAIRE Views
Abstract
Background: A comparative ex vivo study was performed to determine electronic percussive test values (PTVs) measured by cabled and wireless electronic percussive testing (EPT) devices and to evaluate the intra-and interobserver reliability of the wireless EPT device. Methods: Forty implants were inserted into the vertebrae and forty into the pelvis of a steer, a safe distance apart. The implants were all 4.3 mm wide and 13 mm long, from the same manufacturer. PTV of each implant was measured by four different examiners, using both EPT devices, and compared. Additionally, the intra-and interobserver reliability of the wireless EPT device was evaluated. Results: Statistically significant differences (P < 0.05) were observed between PTVs made by the two EPT devices. PTVs measured by the wireless EPT device were significantly higher than the cabled EPT device (P < 0.05), indicating lower implant stability. The intraobserver reliability of the wireless EPT device was evaluated as excellent for the measurements in type II bone and good-to-excellent in type IV bone; interobserver reliability was evaluated as fair-to-good in both bone types. Conclusion: The wireless EPT device gives PTVs higher than the cabled EPT device, indicating lower implant stability, and its inter-and intraobserver reliability is good and acceptable.
Description
Bilmenoglu, Caglar/0000-0002-2580-9899; GECKILI, ONUR/0000-0002-7852-3915; Bilhan, Hakan/0000-0003-1787-3003; Cilingir, Altug/0000-0002-0694-5772
Keywords
Dental implants, dentistry, jaw, edentulous, osseointegration, reproducibility of results, research design
Turkish CoHE Thesis Center URL
Fields of Science
Citation
14
WoS Q
Q1
Scopus Q
Q1
Source
Volume
85
Issue
12
Start Page
1786
End Page
1791