Demir, ErsinKemer, BarisIsildak, IbrahimAboul-Enein, Hassan Y.2024-05-252024-05-25201531573-41101875-672710.2174/15734110116661501281934242-s2.0-84936746957https://doi.org/10.2174/1573411011666150128193424https://hdl.handle.net/20.500.14517/282Demir, Ersin/0000-0001-9180-0609; isildak, ibrahim/0000-0001-9654-1386A new polyvinyl chloride (PVC) membrane Fe3+ - ion selective electrode was constructed as a detector for the analysis Fe3+ using a home-made low-volume flow through cell. Optimum working conditions for the flow injection analysis (FIA) measurements were examined. In addition, under optimum conditions, potentiometric performance characteristics of the Fe3+ ion-selective electrode were investigated. Electrode had a linear response in the concentration range of 1.0 x 10(-3)-1.0 x 10(-6) M (R-2=0.9961, y=24.7+772.0). Reproducibility of the electrode responses was very satisfied. The potential of Fe3+ ion selective electrode was decreased intra and inter day 0.9 % and 3.7% respectively. The designed flow-through cell system as Fe3+ detector revealed sampling rates of approximately 50 injections per hour. The obtained results in FIA were in a good agreement with AAs results at % 95 confidence level for each syrups.eninfo:eu-repo/semantics/closedAccessIron-selective electrodedeterminationflow injection systempotentiometersyrups analysisFe<SUP>3+</SUP> - Ion Selective Electrode Developed as a Detector in Flow Injection AnalysisArticleQ3Q2112104108WOS:000350481300004