A Novel Iron(III)-Selective Membrane Potentiometric Sensor Based on 5-Chloro-3-[4-(trifluoromethoxy) phenylimino] Indolin-2-one
dc.authorid | Demir, Ersin/0000-0001-9180-0609 | |
dc.authorwosid | Demir, Ersin/V-6633-2019 | |
dc.authorwosid | Bekircan, Olcay/AAB-4173-2021 | |
dc.contributor.author | Demir, Ersin | |
dc.contributor.author | Kemer, Baris | |
dc.contributor.author | Bekircan, Olcay | |
dc.contributor.author | Aboul-Enein, Hassan Y. | |
dc.date.accessioned | 2024-10-15T20:19:04Z | |
dc.date.available | 2024-10-15T20:19:04Z | |
dc.date.issued | 2015 | |
dc.department | Okan University | en_US |
dc.department-temp | [Demir, Ersin] Okan Univ, Vocat Sch Hlth Serv, Perfus Tech, TR-34959 Istanbul, Turkey; [Kemer, Baris; Bekircan, Olcay] Karadeniz Tech Univ, Fac Sci, Dept Chem, TR-61100 Trabzon, Turkey; [Aboul-Enein, Hassan Y.] Natl Res Ctr, Pharmaceut & Med Chem Dept, Pharmaceut & Drug Ind Res Div, Cairo 12311, Egypt | en_US |
dc.description | Demir, Ersin/0000-0001-9180-0609 | en_US |
dc.description.abstract | A novel polyvinyl chloride (PVC) membrane sensor that is highly selective to Fe3+ ions was prepared by using 5-chloro-3-[4-(trifluoromethoxy) phenylimino] indolin-2-one (CFMEPI) ionophore. The sensor exhibits a Nernstian response for Fe3+ ions over a wide concentration range (1.0 x10(-2) -1.0x10(-6)M) with a slope of 46.7 +/- 0.5 mV per decade. The sensor has a response time of 20s and can be used for at least 3 months without any measurable divergence in potential. It was concluded that the sensor response was pH independent in the range of 4.0-8.61. The sensor has some advantages such as short analysis time, particularly high selectivity towards iron (III). The sensor was used successfully for direct determination of Fe3+ in several synthetic and real samples with satisfactory results. | en_US |
dc.description.woscitationindex | Science Citation Index Expanded | |
dc.identifier.citation | 11 | |
dc.identifier.doi | [WOS-DOI-BELIRLENECEK-165] | |
dc.identifier.endpage | 35 | en_US |
dc.identifier.issn | 1573-4110 | |
dc.identifier.issn | 1875-6727 | |
dc.identifier.issue | 1 | en_US |
dc.identifier.scopusquality | Q2 | |
dc.identifier.startpage | 29 | en_US |
dc.identifier.uri | https://hdl.handle.net/20.500.14517/6422 | |
dc.identifier.volume | 11 | en_US |
dc.identifier.wos | WOS:000350480400004 | |
dc.identifier.wosquality | Q3 | |
dc.language.iso | en | |
dc.publisher | Bentham Science Publ Ltd | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | Fe3+-selective electrode | en_US |
dc.subject | PVC membrane | en_US |
dc.subject | All-solid-state | en_US |
dc.subject | determination of Fe3+ | en_US |
dc.subject | potentiometer | en_US |
dc.title | A Novel Iron(III)-Selective Membrane Potentiometric Sensor Based on 5-Chloro-3-[4-(trifluoromethoxy) phenylimino] Indolin-2-one | en_US |
dc.type | Article | en_US |
dspace.entity.type | Publication |