Determination of the Optimum Torque to Tighten the Smartpegs of Magnetic Resonance Frequency Analyses Devices: An Ex Vivo Study

dc.authorid Bilmenoglu, Caglar/0000-0002-2580-9899
dc.authorid GECKILI, ONUR/0000-0002-7852-3915
dc.authorid Cilingir, Altug/0000-0002-0694-5772
dc.authorid Bilhan, Hakan/0000-0003-1787-3003
dc.authorscopusid 15135391400
dc.authorscopusid 14012620400
dc.authorscopusid 16548947000
dc.authorscopusid 54880603900
dc.authorscopusid 35267920900
dc.authorwosid BURAL, HAYRUNNİSA CANAN/AAE-1151-2020
dc.authorwosid Bilmenoglu, Caglar/HGE-5706-2022
dc.authorwosid GECKILI, ONUR/AAD-4973-2020
dc.authorwosid Cilingir, Altug/N-2262-2018
dc.contributor.author Geckili, Onur
dc.contributor.author Cilingir, Altug
dc.contributor.author Bural, Canan
dc.contributor.author Bilmenoglu, Caglar
dc.contributor.author Bilhan, Hakan
dc.date.accessioned 2024-05-25T11:18:11Z
dc.date.available 2024-05-25T11:18:11Z
dc.date.issued 2015
dc.department Okan University en_US
dc.department-temp [Geckili, Onur; Bural, Canan] Istanbul Univ, Fac Dent, Dept Prosthodont, Istanbul, Turkey; [Cilingir, Altug; Bilmenoglu, Caglar] Trakya Univ, Fac Dent, Dept Prosthodont, Edirne, Turkey; [Bilhan, Hakan] Okan Univ, Fac Dent, Dept Prosthodont, Istanbul, Turkey en_US
dc.description Bilmenoglu, Caglar/0000-0002-2580-9899; GECKILI, ONUR/0000-0002-7852-3915; Cilingir, Altug/0000-0002-0694-5772; Bilhan, Hakan/0000-0003-1787-3003 en_US
dc.description.abstract The aim of this study was to find an optimal value for tightening the Smartpegs for magnetic radiofrequency analysis devices (RFAs) devices. Thirty implants were placed in 3 cow ribs. The RFA value of each implant was measured in buccal and mesial directions after tightening the Smartpegs with 1, 3, 4, 8, 9, 10, and 11 Ncm. Additionally, 4 different examiners measured the RFA after hand tightening the Smartpegs, and the results were compared. The buccal implant stability quotient (ISQ) values when the Smartpegs were tightened to 1Ncm were significantly lower than the ISQ values when the Smartpegs were tightened to 3, 4, 8, 9, 10, and 11 Ncm (P<.05). The mesial ISQ values when the Smartpegs were tightened to 1, 3, and 4 Ncm were significantly lower than the ISQ values when the Smartpegs were tightened with higher torque values (P<.05). The buccal measurements made by 1 examiner was significantly lower than 3 Ncm (P<.05), and the buccal measurements made by 1 examiner was significantly lower than 4 Ncm (P<.05). The mesial ISQ values measured by 2 examiners were significantly lower than 3 Ncm (P<.05), and the mesial ISQ values measured by the other 2 examiners were significantly lower than 8 Ncm (P<.05). The tightening of the Smartpegs should be standardized by the manufacturer to a range of 5-8 Ncm in order to gain reliable objective RFA values, instead of leaving it to subjective finger pressure. en_US
dc.identifier.citationcount 8
dc.identifier.doi 10.1563/aaid-joi-D-14-00266
dc.identifier.endpage E256 en_US
dc.identifier.issn 0160-6972
dc.identifier.issn 1548-1336
dc.identifier.issue 6 en_US
dc.identifier.pmid 25668680
dc.identifier.scopus 2-s2.0-84949937508
dc.identifier.scopusquality Q2
dc.identifier.startpage E252 en_US
dc.identifier.uri https://doi.org/10.1563/aaid-joi-D-14-00266
dc.identifier.uri https://hdl.handle.net/20.500.14517/308
dc.identifier.volume 41 en_US
dc.identifier.wos WOS:000366723300003
dc.identifier.wosquality Q4
dc.language.iso en
dc.publisher Allen Press inc en_US
dc.relation.publicationcategory Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı en_US
dc.rights info:eu-repo/semantics/openAccess en_US
dc.scopus.citedbyCount 8
dc.subject dental implants en_US
dc.subject radiofrequency analysis en_US
dc.subject osseointegration en_US
dc.subject primary stability en_US
dc.title Determination of the Optimum Torque to Tighten the Smartpegs of Magnetic Resonance Frequency Analyses Devices: An Ex Vivo Study en_US
dc.type Article en_US
dc.wos.citedbyCount 9